beautypg.com

Source[1..n]:stable – EXFO IQS-2400 WDM Laser Source for IQS-500/600 User Manual

Page 162

background image

SCPI Command Reference

156

IQS-2400

Product-Specific Commands—Description

:SOURce[1..n]:STABle?

Description

This query indicates whether the source is stable

or stabilizing.

At *RST, the resolution mode, the wavelength,

the power, the delta temperature and the delta

current are set to the default instrument values.

Syntax

:SOURce[1..n]:STABle?

Parameter(s)

None

Response Syntax

Response(s)

Stable:
The response data syntax for is

defined as a

element.

The response is the stabilization

state.

0- The instrument is stabilizing.

1- The instrument has stabilized.

Example(s)

SOUR:STAB?

Notes

After the mode has been changed, the

instrument needs some time to stabilize. It is

very important not to change the mode while the

instrument is stabilizing. This would cause the

instrument to return to its central wavelength in

Normal mode. Before changing the mode, you

should poll SOUR:STAB? until it returns 1,

indicating that the instrument has stabilized.

See Also

SOURce[1..n]:POWer:STATe?