Source[1..n]:current:delta – EXFO IQS-2400 WDM Laser Source for IQS-500/600 User Manual

Page 103

background image

SCPI Command Reference

WDM Laser Source

97

Product-Specific Commands—Description

:SOURce[1..n]:CURRent:DELTa

Description

This command sets the delta current.

At *RST, the delta current is device-dependent.

Syntax

:SOURce[1..n]:CURRent:DELTa|

MAXimum|MINimum|DEFault

Parameter(s)

Delta:
The program data syntax for is defined

as a element. The

special forms MINimum, MAXimum and DEFault

are accepted on input.

MINimum allows to set the instrument to the

smallest supported value.

MAXimum allows to set the instrument to the

greatest supported value.

DEFault allows the instrument to select a value

for the parameter.

The parameter is a value representing

the delta current in amperes.

Example(s)

SOUR:PULM:STAT ON

SOUR:WAV:MODE HRES

/* Under FDA jurisdiction, a password must be

entered before turning the laser on. */

SOUR:POW:PROT:SPWD "safekey"

SOUR:POW:STAT ON