SCPI Command Reference

90

IQS-2400

Product-Specific Commands—Description

The value used must be equal to or between the

values returned by

SOURce[1..n]:WAVelength:LIMit:LOW? and

SOURce[1..n]:WAVelength:LIMit:HIGH?. The

SOURce[1..n]:WAVelength:LIMit:STEP? query

returns the instrument sensitivity.

³

Power:

The program data syntax for is defined

as a

element followed by an optional

PROGRAM DATA> element. The allowed

elements are:

DBM|W.

The parameter represents a new

power of the signal in watts or dBm. If you do not

specify a unit, the device will use the current

power unit. To select the current power unit, use

the UNIT[1..n]:POWer command.

The value used must be equal to or between the

values returned by

SOURce[1..n]:POWer:LIMit:LOW? and

SOURce[1..n]:POWer:LIMit:HIGH?. The

SOURce[1..n]:POWer:STEP? query returns the

instrument sensitivity.

Example(s)

SOUR:PULM:STAT ON

SOUR:WAV:MODE NORMAL

SOUR:APCO 1547nm ,7.39DBM

:SOURce[1..n]:APCOutput