SCPI Command Reference

84

IQS-2400

Product-Specific Commands—Description

SOUR:STAB? (repeat until the returned value is

not 1)

SOUR:POW 5.0 DBM

CAL:VAL 5.1DBM

Notes

This command must not be used when the

instrument is in High Wavelength Stability. To

determine the appropriate calibration value,

connect the source output to a power meter. The

power read by the power meter is the parameter

that should be entered with the

CALibration[1..n]:VALue command. To return

the device to its original factory calibration, use

the token DEFault.

The value used must not be more than 10 times

greater or less than 10 times lower than the

factory calibration value. An attempt to exceed

these limits will raise a “Parameter out of range”

error.

See Also

UNIT[1..n]:POWer

SOURce[1..n]:WAVelength:MODE

SOURce[1..n]:PULM:STATe

SOURce[1..n]:POWer:PROTection:SetPassWorD
SOURce[1..n]:POWer:[:LEVel][:IMMediate][:AM

PLitude]

:CALibration[1..n]:VALue