Source[1..n]:setpoint – EXFO IQS-2400 WDM Laser Source for IQS-500/600 User Manual
Page 161

SCPI Command Reference
WDM Laser Source
155
Product-Specific Commands—Description
Notes
No modulation is associated with a setpoint.
Therefore, a setpoint created in Normal or
Normal with Dither modulation mode will be
considered Normal (APC) mode. A setpoint
created in either HWS or On/Off modulation
mode will be in HWS (ACC) mode. In turn, a
setpoint in Normal mode can be used in Normal
or Normal with Dither mode. Also, a setpoint in
HWS mode can be used as well in On/Off
modulation mode. The mode must be supplied
because setpoints in Normal and HWS modes
may have the same name. Therefore, the
application will search for the specified setpoint
with the specified mode. If such a setpoint exists,
the application will adjust the wavelength,
power, current step, and temperature step
according to the setpoint. If the setpoint does not
exist, there will be no change.
Modes: If the instrument is in Normal or Normal
with Dither modes and the setpoint is in Normal
mode, there will be no change to the mode of
the instrument. If the setpoint is in HWS mode,
the application will go to HWS mode. If the
instrument is in HWS or On/Off mode and the
setpoint is in HWS mode, there will be no change
in mode. If the setpoint is in Normal mode, the
instrument will go to Normal mode.
See Also
SOURce[1..n]:CURRent:ACCBase
SOURce[1..n]:CURRent:ACCBase?
:SOURce[1..n]:SETPoint