Source[1..n]:pulm:internal: frequency[1|2 – EXFO IQS-2400 WDM Laser Source for IQS-500/600 User Manual
Page 130
SCPI Command Reference
124
IQS-2400
Product-Specific Commands—Description
:SOURce[1..n]:PULM:INTernal:
FREQuency[1|2]
Description
This command is used to set the frequency of the
internal Dither or On/Off modulation signal. If the
instrument is not in Dither or On/Off modulation
mode, this command will have no effect.
To set the frequency of the dither mode, the
FREQuency header suffix [1|2] must be 1.
To set the frequency of the On/Off mode, the
FREQuency header suffix [1|2] must be 2.
At *RST, the dither frequency is 1000 Hz and the
On/Off modulation is 10 Hz.
Syntax
:SOURce[1..n]:PULM:INTernal:FREQuency[1|2]
DEFault
Parameter(s)
Frequency:
The program data syntax for
defined as a
MAXimum and DEFault are accepted on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.
DEFault allows the instrument to select a value
for the
The
new frequency of the internal dither or On/Off
modulation signal in hertz.