YSI 600LS - User Manual User Manual
Page 42
Sondes
Section 2
YSI Incorporated
Environmental Monitoring Systems Operations Manual
2-36
RECOMMENDED VOLUMES OF CALIBRATION REAGENTS
The approximate volumes of the reagents are specified below for both the upright and inverted orientations.
Note that the volume values are only estimates. The actual amount of calibrator solution required will
depend on how many and what type of other probes are installed in you sonde bulkhead.
Table 1A
6820V2-1 and 6920V2-1 Sondes with Standard Calibration Cup
*
Probe to Calibrate
Upright
Inverted
Conductivity
200ml
150ml
pH/ORP
125ml
175ml
ISE
125ml
175ml
All Optical Sensors
50ml
DO NOT CALIBRATE***
Table 1B
6820V2-1 and 6920V2-1 Sondes with Optional Extended Calibration Cup*
Probe to Calibrate
Upright
Inverted
Conductivity
320ml
150ml
pH/ORP
240ml
175ml
ISE
240ml
175ml
All Optical Sensors
225ml
DO NOT CALIBRATE***
Table 2
6820V2-2 and 6920V2-2 Sondes with Extended Calibration Cup
*
Probe to Calibrate
Upright
Inverted
Conductivity
310ml
150ml
pH/ORP
200ml
150ml
ISE
200ml
150ml
All Optical Sensors
225 ml
DO NOT CALIBRATE***
Table 3
600XL and 600XLM Sondes
Probe to Calibrate
Upright
Inverted
Conductivity
50ml
50ml
pH/ORP
25ml
50ml
Table 4
6600V2-2 Sonde with Short Calibration Cup and Long Cup in Parentheses*
Probe to Calibrate
Upright
Inverted
Conductivity
425ml (650ml)
250ml (250ml)
pH/ORP
300ml (500ml)
250ml (250ml))
ISE
300ml (500ml)
250ml (250ml)
All Optical Sensors
180ml (500ml)
DO NOT CALIBRATE***
Table 5
6600EDSV2-2 Sonde with Short Calibration Cup and Long Cup in Parentheses*
Probe to Calibrate
Upright
Inverted
Conductivity
275ml (520ml)
350ml (350ml)
pH/ORP
175ml (400ml)
350ml (350ml)
All Optical Sensors
225ml (420ml)
DO NOT CALIBRATE***