Aquamanager and the timer menu – YSI 5200A User Manual
Page 83
YSI 5200A
165
YSI 5200A
164
Configuring the 5200A
6. Press
for feed timer relay to energize for configured run test time.
Countdown of relay “on” time is displayed.
7. When run time expires, the relay de-energizes. Enter the amount of food
dispensed at the Measured Weight screen.
8. Press
to save settings and return to Run Screen.
Notes:
- Since feeders and food size vary, run several calibrations and enter the
average weight at the last calibration. If feeder or feed type is changed,
recalibrate the feeder.
- If no test/calibration has been performed or if the results have not been
saved, the 5200A will be unable to perform any feeding(s) because there
will be no reference to determine the rate at which feeding(s) will oc-
cur.
- Stored calibration data is available from the AquaManager program.
- If feed timer units change, the feeder must be recalibrated.
- If an incorrect value is entered and saved, another calibration must be
performed.
- Disable the feed timer system when performing the feed timer calibration,
de-energizing any active feed timer relay.
- After calibration is complete, press
to Run Screen. The feed timer
system resets based on the new calibration configuration.
- Feed Timer calibration cannot be done using AquaManager.
AquaManager and the Timer Menu
Most Timer menus can be configured using AquaManager. After polling the 5200A
node, use “Systems Management” to view and upload 5200A configurations. Select
the Timers/Feeder tabs to view and change configuration.
Configuring the 5200A