Timers 1-4, Feed smart, Timer – YSI 5200A User Manual
Page 79

YSI 5200A
157
YSI 5200A
156
am). The timer relay de-energizes at 12:24 pm and not at 12:00 pm which
would have been the original off time if the unit had not been placed in
downloader.
- If power fails when a timer is energized and the power is restored when
the timer system is still active, the timer relay will energize for the timer
duration and not for the remaining original timer cycle. Example - timer
one is configured to start at 10:40 for one hour. The power fails at 11:30.
Power is restored at 11:35. The timer will energize for one hour, not the
remaining 5 minutes of the original timer cycle.
- When the timer relay is energized after the timer duration is changed, the
timer relay de-energizes and reenergizes for the newly configured duration
and not for the remaining original timer duration. Example - timer one
is configured to turn on at 8:30 am for 10 minute duration. Timer one
relay energizes at 8:30 am. At 8:35 am, timer one duration is changed to
20 minute. Timer one relay de-energizes at 8:35 am and re-energizes for
the newly configured 20 minute duration.
- When the timer start time and duration are reconfigured and fall within
the current 5200A unit time, the relay energizes for the entire duration.
Example - The 5200A unit time is 8:30 am. Timer one is configured to
turn on at 8:00 am for one hour. The timer relay will energize at exit to
run screen and will remain energized for one hour. The 5200A does not
use the system clock to calculate duration in this case. The following day,
the relay will energize at 8:00 am, and de-energize at 9:00 am.
- There is a 30 second delay at power up before timers become active.
- Timer relay off times drift by < .2% when duration is longer than 5
minutes. The drift increases the relay ‘on’ time.
Timers 1-4
Menus →Timers→Timer 1-4
Configure up to 40 daily timer events using the 10 events of Daily Timer systems
1-4. Timer events occur on a daily basis until either the start time is cleared, the
timer system is disabled, or the relay is disabled. Timer icon displays at Run Screen
when the timer relay is active. See Icons - page 70.
Configure Timer 1-4:
Configuring the 5200A
1. Install and wire timer devices - page 46.
2. Enter Timer 1-4 menu
Menus →Timers
3.
Use
▲
and
▼
to scroll and highlight submenus.
4.
Press
to select.
5. Enable and configure submenus.
6. Press
to save configuration and return to Run Screen.
Notes:
- Press
to display timer events.
- Timer events with overlapping times occur as follows. Example - event
1 starts at 10:15 for 25 minutes, event 2 starts at 10:25 for 25 minutes;
relay turns on at 10:15, relay turns off at 10:25, relay turns on at 10:25,
relay turns off at 10:50.
- Configure “0” as the duration time to disable any of the 10 timer
events.
- Timers configured for over 86399 seconds (# of seconds in a day) are
not valid.
Feed Smart
tm
Timer
Menus →Timers→Feed Smart
tm
Timer
The 5200A has one feed timer. In order to configure feed timer correctly, it is impor-
tant to understand how the 5200A operates based on the feed timer configuration.
The Feed Smart
tm
Timer system must be calibrated prior to use. For proper opera-
tion, the Feed Timer System requires the user to enter the total amount of food to
be dispensed, number of feedings to dispense the food, and a defined time period
within which the feedings will occur.
The Feed Timer can be set to operate for any or all of the days of the week and
includes continuous or timed feeding modes, FCR (feed conversion ratio), and
sensor options.
Configuring the 5200A