YSI 5200A User Manual
Page 58
YSI 5200A
115
YSI 5200A
114
6. AquaManager Program Folder includes the latest version of instru-
ment firmware. The lastest firmware is stored as an .mdb file in the PC
Program Files>AquaManager>Firmware. At AquaManager, Firmware
Updater>File>Load Firmware Database. Locate and load this file into the
Firmware Updater.
7. Left click
button.
window opens.
8. Select “Yes” and the following screens will appear.
On
instrument
On
PC
YSI 5200A/5400/5500 Bootstrap Loader screen
Flash Programmer uploading firmware into
5200A
Configuring the 5200A
Note:
- If the YSI 5200A/5400 Bootstrap Loader screen does not appear at the
instrument after above steps manually place 5200A in downloader at
Menus → System → Downloader. Close and reopen AquaManager. Reopen
Flash Programmer and repeat steps starting at 6.
- Power cycle to exit downloader and return to run screen. Do not power
cycle during an active upload.
- If after manually placing the 5200A into downloader and the Flash Pro-
grammer utility still does not upload firmware see Force Downloader
- page 116.
9. The 5200A will reboot (power cycle) when firmware upload is complete.
10. Select “ok” when this window appears
11. Close Flash Programmer.
12. Verify version information to confirm downloaded code:
13. Menus → System → Version
14. If applicable, disable Force Serial Port configuration - page 100.
Notes:
- During upload, 5200A systems are placed into Service Hold which de-
energizes any active systems including energized relays. See SVR Hold
Menu - page 169 for additional information. No “off” event is created
for system relays that de-energize when 5200A is placed in downloader.
Systems are reset when power is cycled after downloader.
- All logs and instrument configuration are retained when 5200A firmware
is updated.
- Display board code loads first, followed by control board code. The display
board screen flickers when display board code is loading. The display stops
flickering when the control board code is loading. The Firmware Updater
progress bar will complete two full scrolls during firmware updating.
- Total download time is approximately nine minutes; eight minutes to
Configuring the 5200A