YSI 5200A

115

YSI 5200A

114

6. AquaManager Program Folder includes the latest version of instru-

ment firmware. The lastest firmware is stored as an .mdb file in the PC

Program Files>AquaManager>Firmware. At AquaManager, Firmware

Updater>File>Load Firmware Database. Locate and load this file into the

Firmware Updater.

7. Left click

button.

window opens.

8. Select “Yes” and the following screens will appear.

On

instrument

On

PC

YSI 5200A/5400/5500 Bootstrap Loader screen

Flash Programmer uploading firmware into

5200A

Configuring the 5200A

Note:

- If the YSI 5200A/5400 Bootstrap Loader screen does not appear at the

instrument after above steps manually place 5200A in downloader at

Menus → System → Downloader. Close and reopen AquaManager. Reopen

Flash Programmer and repeat steps starting at 6.

- Power cycle to exit downloader and return to run screen. Do not power

cycle during an active upload.

- If after manually placing the 5200A into downloader and the Flash Pro-

grammer utility still does not upload firmware see Force Downloader

- page 116.

9. The 5200A will reboot (power cycle) when firmware upload is complete.

10. Select “ok” when this window appears

11. Close Flash Programmer.

12. Verify version information to confirm downloaded code:

13. Menus → System → Version

14. If applicable, disable Force Serial Port configuration - page 100.

Notes:

- During upload, 5200A systems are placed into Service Hold which de-

energizes any active systems including energized relays. See SVR Hold

Menu - page 169 for additional information. No “off” event is created

for system relays that de-energize when 5200A is placed in downloader.

Systems are reset when power is cycled after downloader.

- All logs and instrument configuration are retained when 5200A firmware

is updated.

- Display board code loads first, followed by control board code. The display

board screen flickers when display board code is loading. The display stops

flickering when the control board code is loading. The Firmware Updater

progress bar will complete two full scrolls during firmware updating.

- Total download time is approximately nine minutes; eight minutes to

Configuring the 5200A