Ethernet (tcp/ip) module communication, Step 9 connect probe/cable assembly – YSI 5200A User Manual

Page 31

background image

YSI 5200A

61

YSI 5200A

60

D

o not c

onnec

t or disc

onnec

t IDC r

ibbon cable

when 5200A is po

w

er

ed

. S

er

ious damage can oc

cur

.

Figure 3.29

6. Complete other wiring to 5200A.

7. Close front panel - page 40.

8. Test RS232 communication.

Notes:

- When using a USB-Serial Adapter, verify that the port number assigned

during driver installation is the same port configured when using Aqua-

Manager and/or the Firmware Updater.

- Use only a Tripp-Lite Keyspan USB-Serial Adapter, model USA-19HS YSI

p/n

773

(64-bit).

- When making RS232 cable runs greater than 100 feet (30.5 m), use an RS232

Line Extender/Booster. Mount the booster in an area where it is protected

from moisture.

Test RS232 Communications

Apply power to 5200A. Insert node into AquaManager program. Poll node. If

wired correctly, AquaManager will get 5200A configuration and logs.

Installation and Wiring

Ethernet (TCP/IP) Module Communication

The optional serial to ethernet device must be installed on the 5200A control board

when using TCP/IP for communication. Complete installation instructions are

provided with the optional serial to ethernet device.

Notes:

- See Menus → System → Communications → Ethernet beginning on page

90 for configuration information.

- An SMTP server accessible by the local network must be provided by your

local IT support for ethernet connection.

Consult the network administrator to verify compatibility of ethernet configura-

tion.

Step 9 Connect Probe/Cable Assembly

Either a YSI 5561(DO & temp) or YSI 5562 (DO, temp, conductivity, pH & ORP)

probe/cable assembly can be used with the 5200A. Both probe/cable assemblies

have an MS19 connector (Military Spec). Connect the probe/cable assembly as

described below.

Directions to Connect Probe/Cable Assembly

1. Perform steps 1 - 4 of 5200A installation and steps 5 - 8 as required for

the application.

2. Line up pins and guides with the holes and indentations on the cable con-

nector at the bottom of the YSI 5200A - figure 3.30.

3. Hold the cable firmly against the cable connector; then, turn the locking

mechanism clockwise until it snaps into place. (Remove the cable from

the instrument by turning the cable connector counterclockwise until the

cable disengages from the instrument.)

Installation and Wiring