Figure 3.29
6. Complete other wiring to 5200A.
7. Close front panel - page 40.
8. Test RS232 communication.
Notes:
- When using a USB-Serial Adapter, verify that the port number assigned
during driver installation is the same port configured when using Aqua-
Manager and/or the Firmware Updater.
- Use only a Tripp-Lite Keyspan USB-Serial Adapter, model USA-19HS YSI
p/n
773
(64-bit).
- When making RS232 cable runs greater than 100 feet (30.5 m), use an RS232
Line Extender/Booster. Mount the booster in an area where it is protected
from moisture.
Test RS232 Communications
Apply power to 5200A. Insert node into AquaManager program. Poll node. If
wired correctly, AquaManager will get 5200A configuration and logs.
Installation and Wiring
Ethernet (TCP/IP) Module Communication
The optional serial to ethernet device must be installed on the 5200A control board
when using TCP/IP for communication. Complete installation instructions are
provided with the optional serial to ethernet device.
Notes:
- See Menus → System → Communications → Ethernet beginning on page
90 for configuration information.
- An SMTP server accessible by the local network must be provided by your
local IT support for ethernet connection.
Consult the network administrator to verify compatibility of ethernet configura-
tion.
Step 9 Connect Probe/Cable Assembly
Either a YSI 5561(DO & temp) or YSI 5562 (DO, temp, conductivity, pH & ORP)
probe/cable assembly can be used with the 5200A. Both probe/cable assemblies
have an MS19 connector (Military Spec). Connect the probe/cable assembly as
described below.
Directions to Connect Probe/Cable Assembly
1. Perform steps 1 - 4 of 5200A installation and steps 5 - 8 as required for
the application.
2. Line up pins and guides with the holes and indentations on the cable con-
nector at the bottom of the YSI 5200A - figure 3.30.
3. Hold the cable firmly against the cable connector; then, turn the locking
mechanism clockwise until it snaps into place. (Remove the cable from
the instrument by turning the cable connector counterclockwise until the
cable disengages from the instrument.)
Installation and Wiring