Input system specifications – YSI 5200A User Manual
Page 8

YSI 5200A
15
YSI 5200A
14
Input System Specifications
Aux Systems
There are 2 Aux input systems that allow digital or analog (0-
1.0VDC, 0-5.0VDC, or 4-20mA) configuration.
Aux System
Type
Digital
Analog
Range
dry contact open/
close
0 -1.0 VDC, 0-5.0 VDC, or 4-20mA
Accuracy
n/a
0-1.0VDC: ± 3% ± 10 mV
0-5.0VDC: ± 3% ±15 mV
4-20 mA: ± 3% ±100μA
Resolution
n/a
user selectable (0), (0 .0), or (0.00)
Probe assembly 5562 (DO, temperature, pH, ORP, and conductivity sensors)
Probe assembly 5561 (DO and temperature sensors)
Cable grade
military-spec (MS) connector
Cable lengths
4, 10, 20 m (special order up to 100m for DO and temp only)
Temperature
Sensor type
10,000 ohm @ 25°C thermistor
Range
0-45°C (32-113°F)
Accuracy
± 0.2°C
Resolution
0.1°C
Dissolved Oxygen
Sensor type
Membrane covered, steady-state polarographic
Range
0 to 500% air saturation
0 to 60 mg/L
Accuracy
±2% or ± 0.2 mg/L, whichever is greater for 0-200%
±6% of reading for 200-500%
Resolution
0.1% (0.0-99.9%); 1% (> 99.9)
0.01 mg/L (0-10 mg/L); 0.1 mg/L (10-60 mg/L)
Response Time
95% of reading in 17 seconds for 2.0 mil PE membrane;
95% of reading in 36 seconds for 2.0 mil Teflon membrane.
Allow approximately 10-15 minutes after 5200A is powered
on, or the DO sensor is serviced, for the polarographic sensor
to stabilize.
Introduction
Flow requirements minimum flow rate of 3 inches per second or 7.62 cm/second
for 2.0 mil PE membrane; 6 inches per second or 15.24 cm/
second for 2.0 mil Teflon membrane
Other
Automatic salinity compensation for DO readings when a
conductivity sensor is in use and enabled. Manual salinity
compensation for DO readings when a conductivity sensor is
not in use.
pH
Range
0.00 to 14.00 pH
Accuracy
± 0.1 pH
Resolution
0.01 pH
ORP
Range
-2000 mV to +2000 mV
Accuracy
± 20 mV
Resolution
1 mV
Conductivity
Sensor type
4 electrode, Cell constant of 5.0 ±5%
Ranges*
0 to 200 μS
0 to 2000 μS
0 to 20 mS
0 to 200 mS
Accuracy
0 μS to 100 mS, ±0.5% of range
100 mS to 200 mS, ± 1.0% of reading
Resolution
range dependent
0.1 μS (0 to 200 μS)
1.0
μS (0 to 2000 μS)
0.01 mS (0 to 20 mS)
0.1 mS (0 to 200 mS)
Salinity
Parameter Source
1) calculated from temperature & conductivity systems when
enabled
OR
2) user ppt input in DO sensor setup menu
Range
0-80 ppt
Note: When not using a conductivity sensor, the manual
input range is 0-50 ppt.
Introduction