1. Enter Name Devices/Sensors menu
Menus → System → Name Devices/Sensors
2.
Use
▲
and
▼
to scroll and highlight submenus.
3.
Press
to select.
4. Use keypad
to configure name.
5. Press
to save configuration and return to Run Screen.
Notes:
- Instrument, Aux, and Sensor Names are not cleared at Reset to Factory
menus.
- Relay names are cleared at “reset relays”, “reset all systems”, and “reset to
factory/reset all” Service menus.
- The 5200A default system name is displayed at Event, Sensor, and Calibra-
tion logs, regardless if name is user configured.
- Press “Clear” at keypad and then press
to clear user configured
names.
- Aux system names can also be configured at: Menus → System → Con-
figure Aux →Aux 1-2→Probe Name
- Because of font limitations, 1x2 and 2x3 display formats do not show all
20 characters of a 20 character name. Typically only 13 characters are
displayed.
- The instrument name is not shown at Run Screen. The instrument name
is shown at relay enable menus.
- Certain menus (i.e. Configure Aux) display up to 18 characters of the de-
vice name. “...” is shown at the menu if the name exceeds 18 characters.
Password
Menus → System → Password
Enable and set the password in the Password menu. Once the password function
is enabled and saved, the 5200A system menus are not accessible without first
Configuring the 5200A
entering the correct password at the Run Screen. The password can contain up to
12 characters.
1. Enter Password menu
Menus → System → Password
2.
Use
▲
and
▼
to scroll and highlight submenus.
3.
Press
to select.
4. Enable and configure submenus.
5. Press
to save configuration and return to Run Screen.
Notes:
-
is not password protected.
-
icon displays at Run Screen when password function is enabled.
Power Fail Menu
Menus → System → Power Fail
In the event of AC power failure, an AC version instrument’s power switches to
DC if properly wired. The switch from AC to DC power is automatic and does not
require 5200A configuration. Enable Power Fail AC Monitor to generate event when
AC power fails and is restored.
Configuring the 5200A