Service hold, Version, E 129 – YSI 5200A User Manual
Page 65
YSI 5200A
129
YSI 5200A
128
1.
Press
to toggle relay on.
2.
Press
to toggle relay off.
3. Press
to save configuration and return to Run Screen.
Note:
- Timer and control icon(s) remain on the Run Screen until the system is
reset. Example - the timer icon remains at the systemwide location until
the end of the timer cycle that was stopped when the relay test menu was
entered even though that relay is not energized. To reset systems after
exiting the Relay Test menu, power cycle the 5200A.
Service Hold
Menus → System → Service Hold
Pressing
at the Run Screen puts the 5200A into Service Hold. The Service
Hold time is user selectable (1-255 sec, min, or hours). Service Hold turns off any
active relays and disables any sensor, aux, and timer system relays from energizing
regardless of value or condition. 5200A alarm emails are not sent during service
hold. After the service hold time expires or is cancelled, sensor systems are reset
after 30 seconds. Timer systems are not reset. See Service Hold Menu - page 169
for additional information.
Configuring the 5200A
1.
Use
▲
and
▼
to scroll and highlight submenus.
2.
Press
to select.
3. Enable and configure submenus.
4. Press
to save configuration and return to Run Screen.
Note:
- Place the 5200A in service hold as instructed in step 3 of calibration checks
in Maintenance, page183
Version
Menus → System →Version
The Version menu contains information about the 5200A including: display and
control board firmware code version, firmware compile date and time, control board
temperature, AC and DC voltage and model number.
Notes:
- Firmware version information is required when contacting YSI sup-
port.
- The clock chip temperature range starts at 0oC. Values < 0oC are not valid
numbers.
- Firmware date and time compilation are displayed below the version
number.
- The normal operating temperature range will be a few degrees above
ambient.
- AC powered units will run at a higher temperature than DC powered
units.
- To view current AC and DC power values press
to return to the
Run Screen and then return to the menu. Power information does not
refresh when the Version menu is open.
Configuring the 5200A