Directions to wire aux inputs – YSI 5200A User Manual
Page 26
YSI 5200A
51
YSI 5200A
50
Notes:
- All aux inputs have input protection, the maximum input voltage should
not exceed 6 VDC.
- Analog systems will not monitor and/or alarm properly if the device that is
installed and configured operates outside the configured analog range.
- For proper digital input operation, use low impedance switches and wir-
ing. The total resistance should not exceed 200 Ω (ohms).
Directions to Wire Aux Inputs
WARNING: Disconnect external power to the unit before wiring.
AVERTISSEMENT : déconnectez l’alimentation externe de l’unité avant
d’effectuer un câblage quelconque.
1. Perform steps 1 - 4 of 5200A installation.
2. Open front panel - page 39 .
3. Feed aux device wires through drilled hole in rubber grommet to locations
K.
Terminal strip is removable.
CAUTION:
Run high and low voltage cables through separate bulkhead and
conduit.
4. Use (1) or (2) pin to terminate one wire and (G) to terminate ground wire.
(G) pin is the common ground for Aux 1 and 2.
CAUTION
: Aux inputs are not isolated. Devices connected must be electrically
isolated from ground and the water.
Installation and Wiring
D
o not c
onnec
t or disc
onnec
t IDC r
ibbon cable
when 5200A is po
w
er
ed
. S
er
ious damage can oc
cur
.
Figure 3.21
5. Complete other wiring to 5200A.
6. Close front panel - page 40.
7. Test aux systems.
Test Auxiliary Input Devices
Apply power to 5200A. See Aux Setup - beginning on page 51 for configuration
information. Verify display, control, and/or alarm functionality based on user-
defined configuration.
Installation and Wiring