Send test email msg, Email retry delay, Baud rate rs232 – YSI 5200A User Manual
Page 50: E 99
YSI 5200A
99
YSI 5200A
98
- Configuring a 5200A as a Master overrides any slave configuration that
had been previously saved.
- Slave alarms (slave alarm at master and 5200A alarm emails) will not oc-
cur if slave node serial number is not recognized at master node at steps
10. A or 10. B above.
Termination Relay
Menus → System → Communications → Network →Termination Relay
The termination relay must be enabled on the first and last 5200A on a network. All
other devices should have the termination relay disabled.
Send Test Email Msg
Menus → System → Communications → Send Test Email Msg
Send a test email to verify proper configuration of ethernet menus. Test emails are
only transmitted from configured master 5200As.
Note:
- Test emails are sent to all configured email address.
- Exit to Run Screen is required before the test email is sent.
Example of a 5200A test email sent from an ethernet device:
Subject Line: 02/04/10 03:50:05 PM YSI5200A Email Alarm Test Message
From: [email protected]
Sent: Wed, Feb 04, 2010 3:49 pm
Configuring the 5200A
Email Retry Delay
Menus → System → Communications → Email Retry Delay
Configure the Email Retry Delay for the time interval that emails should be retrans-
mitted after the initial alarm email is sent. The initial alarm email is sent immediately
after the general alarm hold off time or after the aux hold off time. Emails will be
retransmitted until one of the bottom front panel softkeys is pressed, regardless if
the alarm condition still exists.
Note:
- On a network, emails are retransmitted according to the slave retry delay
configuration of the slave in alarm and not the master’s email retry delay
configuration. Please note, a slave Email Retry Delay menu will display
(Not available) because of its slave configuration; however, this feature
is available and has a default value of 15 minutes that can be configured
by the user.
- The Retry Delay should be configured at intervals ≥ 15 minutes.
- AquaManager alarm emails are retransmitted based on the 5200A slave(s)
and master email retry delay configuration.
Baud Rate RS232
Menus → System → Communications → Baud Rate
Configure the Baud rate (for RS232 connections only) in the 5200A to match the
baud rate configured in AquaManager. Use the default rate of 115200 for fastest
communication speed.
1. Enter Network menu
Menus → System → Communications → Network →Baud Rate
2.
Press
▲
or
▼
to scroll and highlight a selection.
3.
Press
to confirm selection.
4. Press
to save configuration and return to Run Screen.
Configuring the 5200A