AquaManager and the System Menu
AquaManager users can configure most 5200A system menus from a PC. After poll-
ing the 5200A, use “Systems Management” from the node window to view and upload
5200A configurations. System menus are configured at the Setup and Input tabs.
Configuring the 5200A
Notes:
- Always manually poll the 5200A after uploading configuration changes
and verify the uploaded information is retained. When configuration
changes are uploaded using AquaManager at the same time configura-
tion changes are being made by a user in the menus on the 5200A, the
AquaManager uploads may not be saved depending on the timing of
keystrokes at the 5200A.
- The Run Screen must be displayed during normal operation. If the
5200A is not at the run screen during AquaManager uploading, pressing
at the 5200A will only save certain configuration changes.
See Chapter 5 AquaManager - page 172 and AquaManager online help for ad-
ditional information.
Sensor Setup
Menus → Sensor Setup→Aux 1
→Aux
2
→Conductivity
→DO
→ORP
→pH
→Temperature
→Salinity
Aux, Conductivity, DO, ORP, pH, and temperature system set points, control mode
(set point or PID/PWM), alarm and sensor system relays are configured in the
Sensor Setup menu. In addition, the salinity system is enabled in this menu. Four
onboard relays can be configured for control and alarm output devices. Enabled
relays energize when a control or alarm system is active. Alarm notifications can
be sent via email or cell phone (SMS messaging) - page 95.