Install the weather shield, Step 2 locate and install probe/cable assembly, Location considerations – YSI 5200A User Manual
Page 17
YSI 5200A
33
YSI 5200A
32
8. Place standoffs between 5200A mounting brackets and panel surface -
figure 3.8.
9. Secure 5200A to panel with frame using washers and bolts provided.
Figure 3.8
10. Reattach lanyard by placing lanyard at proper location of display board
panel. Make sure lanyard is fully seated on display panel boss to ensure
proper clearance when front panel is closed - figure 3.7.
11. Replace four screws to reattach display board panel to front panel enclo-
sure - figure 3.6.
12. Connect 5200A display board to ribbon cable -figure 3.5.
13. Secure the 5200A front panel to the frame using mounting screws. Follow
procedures provided at Close Front Panel - page 40 to secure 5200A and
protect internal electronics.
14. Tighten the screws, securing the 5200A to the panel.
Install the Weather Shield
Install the weather shield for outdoor installations. The weather shield can be used
with either a surface or a rail mount. The weather shield kit (model number 6505)
includes a weather shield, u-bolts, pre-drilled metal plate, washers, and nuts.
1. First complete steps 1-2 under Surface Mount or steps 1-3 under Rail
Mount depending on the type of installation. Note: You may need to
modify the pre-drilled holes in the weather shield depending on the
mounting location.
Installation and Wiring
2. At step 3 (surface mount) or step 4 (rail mount), attach the weather shield
to the surface or rail with the metal plate installed to secure the weather
shield to the surface or rail. Figure 3.9 shows the weather shield installed
for a rail mounted 5200A.
3. Tighten the screws, securing the weather shield.
Figure 3.9
Step 2 Locate and Install Probe/Cable Assembly
Use a YSI probe /cable assembly 5561 (with DO and temperature sensors) or a YSI
5562 (with DO, temperature, conductivity, pH, and ORP sensors) on the 5200A.
See Chapter 4 “Configuration”, beginning on page 66, for directions on configur-
ing the 5200A sensor systems. User defined sensor configuration options include:
- Display Formats (i.e. units of measure. For example, mg/L or % saturation,
°C or °F)
- Set point, Control, and Alarm values and ranges
- Control and Alarm relay enabling
After selecting a probe/cable assembly and installing the sensors on the cable, see
Connect Probe Assembly - page 61.
Location Considerations
Measurements taken at the probe location are data logged and processed by the
5200A instrument. Output control and alarm devices are configured to respond
Installation and Wiring