Calibration log, Conductivity, Directions to calibrate conductivity sensor – YSI 5200A User Manual
Page 39
YSI 5200A
77
YSI 5200A
76
Figure 4.5
Notes:
- Insert port plugs into unused sensor port(s) - figure 4.5.
- The sensor must be enabled to calibrate.
- If you use another container for calibration be sure to handle the sensors
will care to protect them from possible physical damage. (See YSI 5562
probe sensor guard - figure 3.12 on page 36 or YSI 5561 probe sensor
guard - figure 3.11 on page 35.) Use a ring stand and clamp to secure
the probe assembly.
- Older sensors may take longer to calibrate than new ones because the sensor
readings may take longer to stabilize.
Configuring the 5200A
Calibration Log
Calibration data is stored in the 5200A calibration log. A calibration data record
includes the type of sensor calibrated, the date and time of the calibration and the
calibration method. Press
at the Run Screen to view the calibration log
- page 168.
Notes:
- Use AquaManager to view and save calibration data on a PC. In addition
to the calibration log information listed above, AquaManager calibration
logs also include user inputs 1 and 2 and the pre and post calibration slopes/
offsets of the sensor(s) when applicable.
- Information from cancelled calibrations is not stored.
Conductivity
Menus → Calibration→ Conductivity
See figure 4.5 to determine which sensor is the conductivity sensor. The conductivity
sensor calibration is a one-point calibration. Calibrate using a traceable conductiv-
ity solution that has a value close to the conductivity value of the water system to
be monitored.
Directions to Calibrate Conductivity Sensor
1. Connect probe/cable assembly - page 61.
2. Read Getting Ready to Calibrate - page 75.
3. Fill calibration/storage cup with 55 ml of traceable conductivity calibration
solution that has a known value.
4. Carefully immerse the sensors into solution. Ensure that the top vent
hole of the conductivity sensor is fully immersed in the solution - figure
4.5, page 76.
5. Gently move the probe module up and down to remove any air bubbles
trapped in the conductivity sensor vent hole.
6. Screw the calibration/storage cup onto the threads of the probe module.
7. Enter conductivity calibration menu:
Menus → Calibration→ Conductivity
8. Allow at least one minute for the temperature to equilibrate.
Configuring the 5200A