Welch Allyn Duet CO2 Module - User Manual User Manual
Page 97

Duet CO
2
Module OEM Implementation Manual
Welch Allyn OEM Technologies
Confidential
Page 97
The Configuration menu provides a means to setup up how the CO
2
Monitor functions. The
configuration items are saved and restored at start up of the CO
2
Monitor application. The
Configuration menu contains the following items:
Configure
Mode
Mainstream
Sidestream
Display
CO2 Module
Units
Sweep Rates
CO2 Scale
Gridlines
Waveform Fill
mmHg
Percent
6.25 mm/sec
12.5 mm/sec
25 mm/sec
50 mmHg
75 mmHg
100 mmHg
14%
10%
6%
Flow Rate
CommPort
90 ml/min
175 ml/min
150 ml/min
Configure\ Mode\ Mainstream (Duet module only) - Checking this enables mainstream
operation and disables sidestream operation.
Configure\ Mode\ Sidestream (Duet module only) - Checking this enables sidestream
operation and disables mainstream operation.
Configure\ Display\Units - The CO
2
Monitor will display CO
2
values in either partial pressure
(mmHg) or percent. These units apply to the waveform and the EtCO
2
and InsCO
2
values.
Configure\ Display\Sweep Rate - The sweep rate setting determines the sweep rate of the CO
2
waveform. This actual sweep rate may vary depending on the PC’s configured display resolution.
Configure\ Display\CO2 Scale - The CO
2
waveform scale can be changed in the
Configure\Display\CO2 Scale menu item. The selections are listed below.
(mmHg)
50 mmHg
(percent)
6%
75 mmHg
10%
100 mmHg
14%
Configure\ Display\Gridlines - Horizontal gridlines on the waveform graph can be enabled by
checking the Gridlines menu item.