Configuration Commands

Configuration commands instruct the module to temporarily modify specific default settings for
measurement criteria. A new pump flow rate may also be specified.

A description of these commands is included in the following table.

Host Command

Description

Change CO

2

waveform update rate

Allows the host to specify how often the module sends a CO

2

waveform data packet. The sensor generates a CO

2

measurement on

every cycle of the chopper motor. The maximum update rate is the
chopper motor’s period of 30 msec. The host may program the rate to
every xx cycles, where xx is the number of cycles per packet.

where xx (ASCII) defines the CO

2

waveform update

rate based on the number of chopper motor cycles.

xx range

00h to FFh

xx default

01h, 30 msec interval

Examples:

send a CO

2

waveform data packet every cycle of

the chopper motor (1

× 30 msec interval)

send a CO

2

waveform data packet every 5

th

cycle of

the chopper motor (5

× 30 msec interval)

send a CO

2

waveform data packet every 255

th

cycle

of the chopper motor (255

× 30 msec interval)

Change breath data clear rate

Allows the host to specify how long the module waits for a new breath
data packet before clearing the breath data (ETCO

2

= 00h / RR = 00h

/ InsCO

2

= 00h).

where xx (ASCII) defines the “no breath” timeout in
sec

xx range

10 to 60 sec (0Ah to 3Ch)

xx default

0Fh, clear breath data after 15 sec of “no breath”

clear the breath data after 10 seconds of “no breath”