beautypg.com

Welch Allyn Duet CO2 Module - User Manual User Manual

Page 38

background image

Duet CO

2

Module OEM Implementation Manual

Page 38

Confidential

Welch Allyn OEM Technologies

Host Command

Module Response

Module Response Format


Request last calibration date

mm = month
dd = day
yyyy = year
Example:

Last cal date = 6/4/2001

Configuration Command Responses

Configuration responses are data packets sent by the module to the host at programmed intervals. A
single host measurement or auto run mode command initiates unsolicited transmission of CO

2

waveform and breath data packets. The programmed interval between packets is either defined by the
host via a configuration command or the host may use the default setting.

Module Response

Module Response Format


Unsolicited CO2 waveform packet at programmed interval

see below


Unsolicited breath data packet at programmed interval

see below

, CO2 Waveform Data

Includes 4 bytes of CO

2

waveform data that are sent at the programmed interval during measurement

and auto run modes.

where xxyy:

xxyy

CO

2

waveform data mmHg

× 256 4*(ASCII)

range:

0 – 99.996 mmHg (0000h – 63FFh)

Example:

CO

2

= 37.50 mmHg

{STX}

“W”

xx

yy

CCITT/CRC

{ETX}

waveform byte string

0x25

0x80

0x79

{STX}W258079{ETX}

02h

57h

32h 35h

38h 30h

37h 39h

03h

CO

2

= 76.65 mmHg

{STX}

“W”

xx

yy

CCITT/CRC

{ETX}

waveform byte string

0x4C

0xA6

0x49

{STX}W4CA649{ETX}

02h

57h

34h 43h

41h 36h

34h 39h

03h