beautypg.com

Welch Allyn Duet CO2 Module - User Manual User Manual

Page 86

background image

Duet CO

2

Module OEM Implementation Manual

Page 86

Confidential

Welch Allyn OEM Technologies

Inclusions

Duet CO

2

Module includes the CO

2

main processor board, mainstream and sidestream daughter

boards, internal and “generic” external pneumatics and the pump. External pneumatics include the
“Water trap receiver to bench” inlet tubing and pump exhaust tubing. “Generic” external pneumatics
include standard lengths of the inlet and exhaust tubing. Custom part numbers must be issued for non-
standard lengths.

MS Duet CO

2

Module includes the CO

2

main processor board and mainstream daughter board.

SS Duet CO

2

Module includes the CO

2

main processor board, sidestream daughter board, internal

and “generic” external pneumatics and pump. Standard 6 in lengths are included for the inlet and
exhaust tubing. Custom part numbers must be issued for non-standard lengths.

Mainstream daughter board does not include the flex circuit. This is typically ordered as service
inventory for replacement of the mainstream daughter board.

Sidestream daughter board does not include the pump, the pneumatic capacitor or the pneumatics
from the Water trap to the sidestream bench inlet. This is typically ordered as service inventory for
replacement of the sidestream daughter board.

Custom mainstream sensor may be tailored to include a custom sensor boot. Details/information/cost
is provided after initial design is approved. Custom part numbers must be issued for custom sensors
and boots.

Flex circuit with Nicolay receptacle may be customized based on OEM requirements.
Details/information/cost is provided after initial design is approved. Custom part numbers must be
issued for non-standard flex circuits.

Inlet and exhaust tubing may be packaged separately by Welch Allyn OEM Technologies or
purchased directly from the manufacturer for service.