Module OEM Implementation Manual
Page 96
Confidential
Welch Allyn OEM Technologies
Menu Options Description
The File menu contains the following items:
Calibration
File
Capture
Simulate
Print Form
Exit
Zero
Two Point
Configure
Enable
Setup
File\Calibration (LC-101 Module only) - Calibration is available under the File menu. Two
different calibrations are available; Zero and Two Point. The Zero calibration is a single point
calibration performed with either room air or reference air with 0% CO
2
. The Two Point
calibration requires a zero reference gas and a span reference gas. The zero can again be
room air or 0% reference gas. The span reference gas must have a CO
2
concentration of
between 8% and 12%, with the exact concentration known to within 0.1%. The span gas
concentration must be manually entered in the Configure menu item.
File\Capture - The Capture feature provide a means of logging CO
2
data and Breath
Parameters to a file. Select the Capture Setup menu item to specify the file name, data
logging rate and what data is to be logged. The data is stored as ASCII text in a CSV format
in the application directory.
File\Simulate - The Simulate feature disables any communications to the CO
2
module and
displays a simulated CO
2
waveform and Breath Parameters. Clicking on this menu item
enables and disables the simulation feature.
File\Print Form - The Print Form selection prints the Main Screen to the default printer.
File\Exit - The Exit selection stops the program.