Module OEM Implementation Manual

Welch Allyn OEM Technologies

Confidential

Page 33

Host Command

Description

Function enable

Allows the host to enable or disable functions listed below.

where xx (ASCII) is the compensation and y (ASCII)
is ON/OFF status.

xx =

00h

O

2

compensation

01h

N

2

0 compensation

02h

desflurane compensation

08h

water vapor compensation

09h

BTPS compensation

0A

Baseline compensation

7F

protect mode

y =

00 binary

OFF

01 binary

ON

3Fh

report current ON/OFF status

Example:

compensate the CO

2

waveform data for N

2

O

turn off protected mode (to allow certain commands)

Change pump flow rate

Allows the host to specify the pump flow rate in milliliters per minute.

where yy (8 bit) is the current pump flow rate.

yy range

90 – 200 ml/min (5Ah – C8h)

yy default

175 ml (AFh)

Example:

set the pump flow rate to 90 ml/minute.

Note: Using this command to set a desired flow rate, the pump flow
rate will revert back to default settings upon power up. The flow rate
can be fixed in the EEPROM to a specific default setting. See
EEPROM Map for details.

Request Software Reset

Allows the host to initiate a software reset.

Note: Duet must be in unprotect mode (See QF70).