Module OEM Implementation Manual
Page 64
Confidential
Welch Allyn OEM Technologies
Adult/Pediatric Multi-Use Airway Adapter Specifications
PHYSICAL SPECIFICATIONS
Overall Dimensions
2.04"L.x 0.88"W x 1.14"H
(5.2cm L x 2.24cm W x 2.9cm H)
Contributory Dead Space
6cc (6ml)
Fittings:
Male O.D. 15mm, per ISO 5356-1 and ASTM F1054
Female I.D. 15mm, per ISO 5356-1 and ASTM F1054
Resistance:
See Figure C
Respiration Rate
0 to 150 BPM
Nominal Weight:
4.8g
Material:
Polysulfone resin, aluminum and sapphire
Pressure Leak Tested at:
15psi
Sterilizing Method:
Steam autoclave
Environmental Conditions
Ambient Shipping/Storage:
Temperature
Relative Humidity
Barometric Pressure
-4° to 140° F (-20° to 60° C)
15-95% RH, non-condensing
795mmHg to 375mmHg (-1300ft to 45000ft)
Ambient Operating Conditions:
Temperature
Relative Humidity
Barometric Pressure
50° to 140° F (10° to 60° C)
15%-95%RH, non-condensing
795mmHg to 375mmHg (-1300ft to 45000ft)
Note:
This airway adapter is designed for a maximum flow of 120 l/min. This airway adapter is
designed to be sterilized via steam autoclave only. Do not exceed 250° F (120° C)
(14.7PSIG). Designed to withstand a minimum of 20 sterilization cycles.
Pressure Drop
(cm H
2
0)
0
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
0
10
20
30
40
50
60
70
80
90
100
110 120
Flow (Liters/min)
Figure C
Average Pressure Drop Across the Adult Airway Adapter