Module OEM Implementation Manual

Page 80

Confidential

Welch Allyn OEM Technologies

14h, CO

2

Sensor Signal Very Low

IR detector voltage measured during the reference gas cell phase of the chopper wheel
rotation is below the preset threshold. Transmission of CO

2

waveform and breath data packets

is halted until the advisory is cleared. The Duet CO

2

Module remains in the measurement

mode despite the condition.

Action Following instructions for 13h fault code.

50h, CO

2

Sensor Disconnect

Duet CO

2

Module does not detect an attached CO

2

sensor.

Action If mainstream operation, try a different sensor. If the fault persists when a CO

2

sensor

is connected, check if the sensor connector is loose by reseating the connector.

51h, CO

2

Sensor Manufacturer ID Mismatch (mainstream only)

Mainstream sensor’s ID stored in the EEPROM does not match the correct OEM specific
code.

Action If mainstream operation, try a different sensor and issue a new measurement mode
command.

60h, CO

2

Sensor EEPROM Error/Revision #

Revision # read from the CO

2

sensor EEPROM is invalid.

Action Check that the revision data is in the 1.xx format. If mainstream operation, try a
different sensor and issue a new measurement mode command.

61h, CO

2

Sensor EEPROM Error/Checksum

Checksum read from the CO

2

sensor EEPROM is invalid.

Action If mainstream operation, try a different sensor and issue a new measurement mode
command.

65h, CO

2

Sensor EEPROM Error/Read/Write

A read or write error to the CO

2

sensor EEPROM has occurred.

Action If mainstream operation, try a different sensor and issue a new measurement mode
command.

66h, CO

2

Sensor EEPROM Error/CRC Code

CRC code read from the CO

2

sensor EEPROM is invalid.

Action If mainstream operation, try a different sensor and issue a new measurement mode
command.