beautypg.com

Welch Allyn Duet CO2 Module - User Manual User Manual

Page 37

background image

Duet CO

2

Module OEM Implementation Manual

Welch Allyn OEM Technologies

Confidential

Page 37

Host Command

Module Response

Module Response Format


Request software version

xxx = version #
mm = version month
dd = version day
yyyy = version year
Example:

version = 1.30
date = 10-23-1998


Request hardware version

xx = hardware version
Example:

hardware version = 2.5


Request ambient barometric
pressure reading from
absolute pressure transducer

xxxx = ambient barometric pressure in
mmHg
Example:

ambient baro pressure =745mmHg


Request sensor temperature

xx = (temperature

× 4) in

o

C

Example:
temperature = (168/4) = 42°C


Request chopper motor speed

xxxx = (motor period

× 10) in msec

Example:

chopper motor speed = 30.0 msec


Request single CO

2

measurement

xx = CO

2

in mmHg

× 256

CO

2

= 16 bit binary integer

Example:

CO2 = 38.80 mmHg


Request single breath packet

xx = ETCO

2

in mmHg

yy = RR in bpm
ww = InsCO

2

in mmHg

Example:

ETCO2 = 37mmHg, RR = 24 bpm,
InsCO2 = 0


Request current pump flow
rate

xxxx = pump flow rate in ml/min
Example:

Flow rate = 175ml/minute


Request sensor EEPROM
revision #

xx = revision #
Example:

revision # = 2


Request sensor OEM specific
code

xx = OEM specific code
Example:

OEM specific code = 00


Request sensor serial #

xx = serial #
Example:
serial # = 13506