Module OEM Implementation Manual

Page 52

Confidential

Welch Allyn OEM Technologies

Requirements

Electromagnetic Interference

The Duet Module meets the following EMC requirements without enclosure.

§ Radiated Emissions - tested per CISPR 11, 1990

§ Magnetic Field Emissions - tested per RE101

Electromagnetic Susceptibility

The Duet Module meets the following EMC requirements without enclosure.

§ Electrostatic Discharge - tested per IEC 1000-4-2, 1995
§ Fast Transient Bursts - tested per IEC 1000-4-4 (EN61000-4-4,1995)
§ Radiated Electromagnetic Fields - tested per IEC 1000-4-3 with dwell time adjusted to 3

seconds (80 MHz - 1000 MHz)

§ Radiated Electromagnetic Fields - tested per IEC 1000-4-3 with dwell time adjusted to 3

seconds (26 MHz - 80 MHz)

§ Conducted Electromagnetic Energy - tested per CS114 (MIL-STD-462D)
§ Magnetic Fields - RS101 (MIL-STD-462D)


Compliance Testing Summary

The Duet Module meets the specifications put forth in the FDA Reviewer Guidance for pre-market
notification submissions of Anesthesiology and Respiratory Devices Branch, November 1993 and IEC
601-1-2, Second Edition, Draft 1, 1996 (EN60601-1-2). The test sample was evaluated in accordance
with specifications put forth in the following:

§ CISPR 11, 1990 (EN 55011, 1991)
§ IEC 1000-4-2, 1992 (EN 61000-4-2, 1995)
§ IEC 1000-4-3, 1995 (ENV 50204)
§ IEC 1000-4-4, 1995 (EN 61000-4-4, 1995)
§ MIL-STD-461D, 11 January 1993
§ RE101, RS101, CS114


ISO 9918 and EN 864

The Duet Module, when implemented according to this manual, meets or exceeds the requirements
outlined in Section 8, Clauses 50.3 through 50.9 and Section 11, Clauses 60 and 61 of the ISO 9918
Standard and equivalent sections of EN 864.