Welch Allyn Duet CO2 Module - User Manual User Manual
Page 32

Duet CO
2
Module OEM Implementation Manual
Page 32
Confidential
Welch Allyn OEM Technologies
Host Command
Description
Change breath data update rate
Allows the host to specify how often the module sends a breath data
packet.
where xx (ASCII) is converted to a 8 bit binary
number (yyzz zzzz) with yy defining how often the
breath data is updated.
If yy = 00 binary, Breath data packet update rate is defined by zzzzzz,
where zzzzzz is in tenths of seconds. The six lower
bits = (seconds x 10). Odd values are rounded up
to even values.
If yy = 01 binary, Breath data is sent on every breath (zzzzzz =
doesn’t matter).
Exception: If the “no breath” condition occurs, data is
sent at a rate defined by the “no breath timeout”
command (
If yy = 10 binary, Breath data is sent only when the data changes
(zzzzzz = doesn’t matter).
Exception: If after 15 seconds the breath data has not
changed, data is sent after the next breath. This
does not guarantee a 15 second update. However,
if the “no breath” condition has occurred, the “no
breath” timeout timer takes precedence.
Default is yy = 10 or xx = 80
If yy = 11 binary, Not defined.
send a breath data packet at .6 sec interval (odd
values rounded up0
send breath data packet every breath
send breath data packet every the data changes
Special case:
no breath data packet sent (Note: Continuous CO
2
mode would use this command.)