O/Desflurane Compensation

The presence of oxygen (O

2

), nitrous oxide (N

2

O), and desflurane at elevated concentrations affects

the measurement of CO

2

concentration. O

2

, N

2

O and desflurane compensations correct the ETCO

2

value during the presence of these interfering agents at high levels

The Duet module provides mathematical compensations that utilize midpoint correction factors for
interfering agents such as Oxygen (O

2

), Nitrous Oxide (N

2

O) and desflurane. The compensations may

be manually enabled by the user to compensate the ETCO

2

value during the presence of these

interfering agents at high levels. As the concentrations of interfering agents move away from the
midpoint values, the correction error becomes greater.

The variables and correction equations are summarized in this table.

O

2

Correction

N

2

O/Desflurane Correction

Operating Conditions

OFF

OFF

O

2

≤ 50%, no N

2

O

ON

OFF

O

2

> 50%, no N

2

O

OFF

ON

O

2

≤ 50% and N

2

O or desflurane

≥ 12%

ON

ON

O

2

> 50% and N

2

O or desflurane

≥ 12%

The O

2

/N

2

O/desflurane compensations are enabled/disabled by the following Function Enable/Disable

commands.

Compensation

Function Enable Command

O

2

compensation OFF

O

2

compensation ON

N

2

O compensation OFF

N

2

O compensation ON

Desflurane compensation OFF

Desflurane compensation ON