Performance Validation

Introduction

Welch Allyn OEM Technologies recommends that the OEM perform annual verification procedures on
the Mainstream and the Sidestream CO

2

sensor and pneumatic system components. Typically,

calibrated gas measurement is a steady state measurement, so updated breath data is not necessary.
Also, calibrated gasses are dry gas, therefore the water vapor and BTPS compensations should be
disabled by the Host in those instances where dry gas readings are made.

Verification Procedures

Mainstream CO

2

Sensor Verification

Equipment needed

§ calibrated Low Dead Space Airway Adapter

§ gas valve and tubing

§ calibration gas canister / 10.00% CO

2

, balance nitrogen, ±0.02%

Illustration of gas canister,
regulator and mainstream circuit

10%
CO2

Bal. N2

Regulator

On/Off

Free Flow
Away From
Monitor Input

Mainstream
Sensor

Airway
Adapter