Module OEM Implementation Manual

Page 54

Confidential

Welch Allyn OEM Technologies

To check the mainstream sensor

• Locate the monitor in a well ventilated room.
• Turn the monitor ON and enter the mainstream mode.
• Disable the Water Vapor and BTPS Compensations, and the Breath Data Packets.
• Warm the mainstream sensor for a minimum of 15 minutes.
• At the end of the 15 minute warm up period, verify that the displayed CO

2

reading is between 0

mmHg and 2 mmHg.

• Connect the mainstream sensor to the airway adapter.
• Connect the calibration gas and apply the gas. The gas flow rate should be set to

approximately 250ml/min.

• Verify that the accuracy of the CO

2

reading is within the manufacturer’s specification for the

calibration gas used.

To troubleshoot error

• Check to make sure that the room is well ventilated. CO

2

readings may be elevated in a closed

room.

• Check that the calibration gas is of known concentration. Repeat the calibration check using a

different mainstream sensor. If the CO

2

readings are consistently elevated or depressed, the

calibration gas may be suspect. Access another gas source.

• Insure that the Mainstream sensor was warmed to the proper operating temperature of 42

o

C.

Repeat the warm up period of 15 minutes and calibration check.

• Check for the presence of a dirty or faulty calibrated airway adapter. Clean the airway adapter

and repeat the calibration check. If the CO

2

reading is still in error, replace the airway adapter

and repeat the calibration check. If different readings persist, a faulty airway adapter is
suspected.

• Check for air leaks in the pneumatic tubing and connections. Repair or replace as needed.