To troubleshoot error

• Make sure that the room is well ventilated. CO

2

readings may be elevated in a closed room.

• Check that the calibration gas is of known concentration. Repeat the calibration check using a

different mainstream sensor. If the CO

2

readings are consistently elevated or depressed, the

calibration gas may be suspect. Access another gas source.

• Insure that the Mainstream sensor was warmed to the proper operating temperature of 42

o

C.

Repeat the warm up period of 15 minutes and calibration check.

• Check for the presence of a dirty or faulty airway adapter. Clean the airway adapter and repeat

the calibration check. If the CO

2

reading is still in error, replace the airway adapter and repeat

the calibration check. If different readings persist, a faulty airway adapter is suspected.

• Check for air leaks in the pneumatic tubing and connections. Repair or replace as needed.

Pneumatic System Verification (Sidestream)

Equipment includes

§ Sidestream Duet CO

2

Module

§ calibrated mass flow meter

(0 – 250 cc/min minimum range / full range accuracy 10% or better)

§ Water trap receiver assembly and Water trap

§ sample line

To perform the pneumatic validation procedure

• Insert the Water trap into the receiver assembly and connect the sample line to the Water trap

inlet.

• Connect the free end of the sample line to the vacuum side of the calibrated flow meter.
• Connect the Sidestream inlet tube to the exhaust end of the receiver assembly.
• Set the monitor to the highest flow rate i.e. 175ml/min.
• Verify that the accuracy of the flow rate reading is within the manufacturer’s specification for

the flow meter used.

• If more than one flow rate is available, repeat the above two steps for each additional flow rate.