MAINTENANCE
10-2
Figure 10-2. UV EPI Lamp Partially Removed
3.
Place the lamp assembly upside down on a flat surface and then rotate each fluorescent tube
90 degrees. Once the metal pins are visible, lift the tubes out and dispose of them.
Figure 10-3. UV EPI Lamp Replacing the Fluorescent Tubes
4.
Reverse the procedure to install the four new fluorescent tubes.
10.3 FUSES
To replace a fuse:
1.
Turn the power switch off and unplug the spot cutter from its power source.
2.
Squeeze the fuse door tabs and pull it open (see Figure 10-4).
3.
Remove the old fuse and then insert a new Type T 2.5A fuse
4.
Replace the fuse door, reconnect the power and turn the instrument on.
Note: If the instrument does not power up, make sure the fuse was installed correctly. In the event the new
fuse has blown or the instrument will not power up, contact your Bio-Rad technical representative
immediately.