EXQUEST CALIBRATION WIZARD

3-5

Figure 3-6. “Head Height Calibratioin” screen.

7.

Flat Field Calibration. The flat field calibration step corrects for inhomogeneity in the light source.
Before you start, make sure the gel tray is installed and that it is clean and dry. Click Next to
proceed with the calibration, Back to repeat the previous step or Cancel to abort the calibration.

Figure 3-7. “Flat Field Calibration” screen.