3.2 Remote control commands
719 S Titrino
93
ON means that the Titrino reports automatically the moment the correspond-
ing change occurs.
.Status:
Global switch for all set AutoInfo.
.P
PowerOn: Simulation of power on (3.2.2.87). Not from mains.
Messages from node .T, Titrator:
.T.R
Ready: Status 'Ready' has been reached.
.T.G
Go: Instrument has been started.
.T.GC
GoCommand: Instrument has received a go command.
.T.S
Stop: Status 'Stop' has been reached.
.T.B
Begin of sequence.
.T.F
Final: End of determination, the final steps will be carried out.
.T.E
Error. Message together with error number, see page 60ff.
.T.H
Hold: Status 'Hold' has been reached.
.T.C
Continue: Continue after hold.
.T.O
Conditioning OK: EP reached (in SET with conditioning).
.T.N
Conditioning Not OK: EP not reached (in SET with conditioning).
.T.Re
Request: In the inquiry of an identification or the sample size after
start of titration.
.T.Si
SiloEmpty: Silo empty, i.e. the last line has been removed from
the silo memory.
.T.EP
EPList: Entry into EP list (with SET)
.T.RC
Results have been recalculated.
Messages for changings in the I/O lines. If the changings are made simulta-
neously, there is 1 message. Pulses receive 2 messages: one message each
for line active and inactive.
.I
Input: Change of an input line.
.O
Output: Change of an output line (except 7, pin 2, for recorder
pulses).
If a change occurs that requires a message, the Titrino sends space (ASCII
32) and ! as an introducer. This is followed by the name of the device (see
3.2.2.47). Special ASCII characters in the device name are ignored. If no de-
vice name has been entered, only ! is sent. Finally the Titrino sends the infor-
mation which node has triggered the message.
Example: !John".T.Si": The message was triggered from instrument "John",
node .T.Si
3.2.2.87. Setup.PowerOn $G
Simulation of 'power on'. The device has the same status as after power on:
The cylinder is filled, error messages deleted and the current sample number
set to 0. The method last used is ready for operation.
3.2.2.88. Setup.Initialise $G
Setup.Initialise.Select ActMeth, Silo, Calib, Config,
Assembly, Setup, All
Setting of default values for the following areas:
ActMeth: Current method. Parameters, calculations, and assignments for
the data output, operands C01...C19.
Silo:
The silo memory is deleted. Same function as delete entire silo.
Calib:
pH calibration data for all measuring inputs.
Config:
All values under &Config.
Assembly: All values under &Assembly.