11 data output – Metrohm 719 S Titrino User Manual
Page 48
2.11. Data output
719 S Titrino
44
2.11 Data output
Reports for the output at the end of a determination
With
determination is defined.
The entries are specific to the method and are stored in
the method memory.
report:
Definition of report blocks to be printed automatically at
the end of the determination.
The display texts of the Titrino are shown to the left.
The values are the default values.
>report
report:
report:full
Report sequence: full, short, param, calc, calib, ff
Select a block with
than one report block, set a ";" as a delimiter between
the blocks.
Meaning of the report blocks:
full
Full result report with raw results, calculations and statistics.
short
Short result report with calculations and statistics.
param Parameter
report.
calc
Report with formulas and operands.
calib Calibration
data.
ff
Form feed on printer.
Original reports which are put out automatically at the end of the titration can be printed
with recalculated values at any time. Key sequence:
Original reports have double dashes ==== at the end, whereas recalculations are
marked by single dashes
----
.
Report outputs can be stopped with
DEF
(
2
def
>report