3.2. Remote control commands
719 S Titrino
68
&Config
Object Description
Input
range
Reference
& Root
:
├
Config
Instrument configuration
├
.PeriphUnit
Selection of peripheral units
│ ├
.CharSet External
printer
Epson,Seiko,Citizen
│ │
IBM,HP 3.2.2.40.
│ ├
.Balance
Selection of balance
Sartorius,Mettler,Mettler AT
│ │
AND,Precisa 3.2.2.41.
│ ├
.Plot
Selection of plot at analog output
U, dU/dt, V, dV/dt
│ │
U(rel), T
ditto
│
├
.Aux
Miscellaneous
│ ├
.Language Dialog
language
english, deutsch,
│ │
francais, espanol, italiano,
│ │
portugese, svenska
3.2.2.42.
│ ├
.Set
Setting of date and time
$G
3.2.2.43.
│ │ ├
.Date Date
XXXX-XX-XX
│ │ ├
.Time Time
XX:XX
│ ├
.RunNo Run
number
0...9999 3.2.2.44.
│ ├
.AutoStart Automatic
start
1...9999,
OFF
3.2.2.45.
│ ├
.StartDelay
Start delay time
0...999 999
3.2.2.46.
│ ├
.DevName
Device label
8 ASCII char.
3.2.2.47.
│ ├
.Prog
Program version
read only
3.2.2.48.
│
├
.RSSet
Settings RS232
$G 3.2.2.49.
│ ├
.Baud Baud
rate
300,600,1200,2400,4800,
│
│
9600
│ ├
.DataBit
Number of data bits
7, 8 ditto
│ ├
.StopBit
Number of stop bits
1, 2
ditto
│ ├
.Parity Parity
even,
odd,
none
ditto
│ ├
.Handsh Handshake
HWs, HWf, SWchar,
│
│
SWline, none
ditto
│
├
.ComVar
Values of common variables
│ ├
.C30 C30
0... ±999 999
3.2.2.50.
│ ├
up to C39
0... ±999 999
│