2.4.2 Development of a method, titration of an acid

Selection of the mode

4 × repeatedly until "pH" appears again in

the display.

Confirm the measured quantity "pH" with

.

Now set the end point.

Entry of the end point and the control parameters

• Press . The display shows:

Control parameters for EP1


• Set the end point to pH = 7.00.

• Dynamics means the control range in which the

adding of the titrant is controlled in dependence on

the measured value (OFF: largest control range, i.e.

slow titration).

Set the range to 3. I.e. 3 pH units before the set EP

controlled dosing starts.

Maximum possible titration rate inside the control

range.

mode

mode SET

SET pH

********

SET: pH

SET pH

********

parameters

>SET1

>SET1

EP at pH

OFF

>SET1

EP at pH

7.00

>SET1

dynamics OFF

>SET1

dynamics 3

>SET1

max.rate 10.0

ml/min