2 development of a method, titration of an acid – Metrohm 719 S Titrino User Manual
Page 15
2.4
Tutorial
719 S Titrino
11
2.4.2 Development of a method, titration of an acid
Selection of the mode
4 ×
• Press
the display. For a description of the SET mode see
page 30ff.
• Confirm "SET" with
• Now select the measured quantity: Press
the display.
Confirm the measured quantity "pH" with
Now set the end point.
Entry of the end point and the control parameters
• Press . The display shows:
Control parameters for EP1
• Set the end point to pH = 7.00.
• Dynamics means the control range in which the
adding of the titrant is controlled in dependence on
the measured value (OFF: largest control range, i.e.
slow titration).
Set the range to 3. I.e. 3 pH units before the set EP
controlled dosing starts.
Maximum possible titration rate inside the control
range.
mode
mode SET
SET pH
********
SET: pH
SET pH
********
parameters
>SET1
>SET1
EP at pH
OFF
>SET1
EP at pH
7.00
>SET1
dynamics OFF
>SET1
dynamics 3
>SET1
max.rate 10.0
ml/min