2.7 Parameters, key
719 S Titrino
35
electrode test: OFF
temperature 25.0 °C
Electrode test (OFF, ON)
Test for polarized electrodes. Performed on
changeover from the inactive standby mode to a
measurement. "OFF" means that the test is not
performed.
Temperature (-170.0...500.0 °C)
Temperature is measured at the start if a T sensor is
connected.
This parameter is used for temperature compensation
in pH measurements.
>statistics
status: OFF
mean n= 2
res.tab: original
delete n= 1
Statistics calculation
Status of statistics calculation (OFF, ON)
If the statistics calculation is switched off, the following
inquiries regarding the statistics do not appear.
Mean value calculation from n single results (2...20)
Result table (original, delete n, delete all)
"original":
The original table is used. Deleted
individual results are again incorporated in
the evaluation.
"delete n": Deletion of single results with the index n.
"delete all": The entire table is deleted.
Delete data from sample number n (1...20)
The deleted result is removed from the statistics
calculation.
>preselections
req.ident: OFF
req.smpl size: OFF
activate pulse: OFF
Preselections for the sequence
Request of sample identifications after start of titration
(id1, id1&2, all, OFF)
After start, sample identifications can be inquired
automatically: Only id1, id1 & id2, all three id's or no
inquiries.
Request of sample size after start of titration (value, unit,
all, OFF)
Pulse output on line "activate" (L6, pin 1) of the remote
socket (ON, OFF)
see page 131.