3.1.3 Status messages
In order to have an efficient control by an external control device, it must also
be possible to query status conditions; they provide information on the status
of the Titrino. The trigger $D initiates output of the status. Status messages
consist of the global status, the detailed status and eventual error messages,
e.g. $S.Mode.SET;E26. The global status informs on the activity of the proc-
ess, while the detailed status conditions show the exact activity within the
process.
The following global status conditions are possible:
$G
Go:
The Titrino is executing the last command.
$H
Hold:
The Titrino has been held ($H, key
the hold status)
$C
Continue:
The Titrino has been restarted actively after hold
$R
Ready:
The Titrino has executed the last command and is ready
$S
Stop:
A process has been aborted in an "unnatural manner". e.g. stopped or aborted
because there was an error.
Detailed status conditions
Status conditions of the global $G:
$G .Mode.SET .Inac:
Instrument at the beginning or at the end of a titration.
.Req
.Id1:
Instrument in the SET mode, requesting Id1 after start.
.Id2:
Instrument in the SET mode, requesting Id2 after start.
.Id3:
Instrument in the SET mode, requesting Id3 after start.
.Smpl:
Instrument in the SET mode, requesting sample size after start.
.Unit:
Instrument in the SET mode, requesting unit of sample size after start.
.Start:
Instrument in the SET mode, processing the start conditions.
.SET1:
Instrument in the SET mode, titrating to the first endpoint.
.SET2:
Instrument in the SET mode, titrating to the second endpoint.
.Cond.Ok:
Instrument in the SET, conditioning, endpoint reached (after the first
startup from the standby mode).
.Cond.Prog:
Instrument in the SET mode, conditioning, endpoint not reached
(Conditioning progressing).
$G .Mode.MEAS .Inac:
Instrument at the beginning or at the end of a titration.
.Req .Id1:
Instrument in the MEAS mode, requesting Id1 after start.
.Id2:
Instrument in the MEAS mode, requesting Id2 after start.
.Id3:
Instrument in the MEAS mode, requesting Id3 after start.
.Smpl:
Instrument in the MEAS mode, requesting sample size after start.
.Unit:
Instrument in the MEAS mode, requesting unit of sample size after start.
.Meas:
Instrument in the MEAS mode, measuring.
$G .Mode.CAL .Inac:
Instrument at the beginning or at the end of a calibration
.Req.Temp:
Instrument in the CAL mode, requesting calibration temperature.
.Meas.Temp:
Instrument in the CAL mode, measuring calibration temperature.
.Req.Buf1:
Instrument in the CAL mode, requesting pH of buffer 1.
.Meas.Buf1:
Instrument in the CAL mode, measures buffer 1.
.Req.Buf2:
Instrument in the CAL mode, requesting pH of buffer 2.
.Meas.Buf2:
Instrument in the CAL mode, measures buffer 2.
etc.
$G .Assembly.Bur .Fill:
Buret in filling process
.ModeDis:
Buret in DIS mode
Status conditions of the global $H:
The status message of the action which has been held appears.
If the process is held because a monitored limit has been violated, its status
message is $H.Mode.XXX.XXX.