Metrohm 719 S Titrino User Manual
Page 96
3.2. Remote control commands
719 S Titrino
92
Selection of the values from Assembly for the output in the set time interval
(see 3.2.2.82):
.CyclNo: Cycle number of the potential measurement. Together with the
cycle time (3.2.2.81), a time frame can be set up.
The cycle number is set to 0 on switching on the instrument and it
is always incremented as long as the instrument remains
switched on.
.V: Volume
.Meas:
Measured value associated to the cycle number.
The unit "assembly" must be preset (see 3.2.2.83).
3.2.2.85. Setup.SendMeas.Titrator.CyclNo ON, OFF
Setup.SendMeas.Titrator.V ON, OFF
Setup.SendMeas.Titrator.Meas ON, OFF
Setup.SendMeas.Titrator.dVdt ON, OFF
Setup.SendMeas.Titrator.dMeasdt ON, OFF
Setup.SendMeas.Titrator.dMeasdV ON, OFF
Selection of the values from the titrator which are sent in the set time interval
(see 3.2.2.82, formats see 3.2.2.68):
.CyclNo: Cycle number. Together with the cycle time (3.2.2.71), a time
frame can be set up. The other data belong to the corresponding
cycle number. The cycle number is set to 0 at the start of a
method and it is incremented until the end of the method.
.V: Volume.
.Meas: Measuring
value
.dVdt:
associated volume drift.
.dMeasdt: associated measured value drift.
.dMeasdV: associated 1st derivative of the titration curve.
The unit "titrator" must be preset (see 3.2.2.83).
3.2.2.86. Setup.AutoInfo.Status ON, OFF
Setup.AutoInfo.P ON, OFF
Setup.AutoInfo.T.R ON, OFF
Setup.AutoInfo.T.G ON, OFF
Setup.AutoInfo.T.GC ON, OFF
Setup.AutoInfo.T.S ON, OFF
Setup.AutoInfo.T.B ON, OFF
Setup.AutoInfo.T.F ON, OFF
Setup.AutoInfo.T.E ON, OFF
Setup.AutoInfo.T.H ON, OFF
Setup.AutoInfo.T.C ON, OFF
Setup.AutoInfo.T.O ON, OFF
Setup.AutoInfo.T.N ON, OFF
Setup.AutoInfo.T.Re ON, OFF
Setup.AutoInfo.T.Si ON, OFF
Setup.AutoInfo.T.M ON, OFF
Setup.AutoInfo.T.EP ON, OFF
Setup.AutoInfo.T.RC ON, OFF
Setup.AutoInfo.I ON, OFF
Setup.AutoInfo.O ON, OFF