2.5 Configuration, key
719 S Titrino
21
2.5 Configuration, key
Key
The set values apply to all modes.
peripheral units:
Selection of printer, balance and the curve at the ana-
log output.
auxiliaries:
e.g. setting of dialog language, date, time, etc.
RS232 settings:
RS parameters for the COM interface.
common variables:
Values of common variables.
The display texts of the Titrino are shown to the left.
The values are the default values.
>peripheral units
send to: IBM
balance: Sartorius
record: U
Settings for peripheral units
Selection of printer (Epson, Seiko, Citizen, HP, IBM) at
the Titrino Rs232 interface.
"Epson", for Epson
"Seiko", e.g. for DPU-414
"Citizen", e.g. for iDP 562 RS, Custom DP40-S4N
"HP" e.g. for Desk Jet types. Place curves always at the
beginning of a page as you cannot have them over 2
pages.
"IBM" for all printers with IBM character set Table 437
and IBM graphics, as well as for the data transmission
to a computer with Metrodata software TiNET or
VESUV.
Selection of balance (Sartorius, Mettler, Mettler AT,
AND, Precisa)
Sartorius: Models MP8, MC1
Mettler:
Models AM, PM, AX, MX, UMX and
balances with 011, 012, or 016 interface
Mettler AT: Model AT
AND:
Models ER-60, 120, 180, 182, FR-200, 300
and FX-200, 300, 320
Precisa:
Models with RS232C interface
Selection of the curve for the output at the analog
output (U, dU/dt, V, dV/dt, U(rel), T
U:. Voltage
dU/dt: Measured value drift
V: Volume
dV/dt Volume drift
U(rel): Control deviation with SET
T: Temperature with MEAS T
CONFIG
configuration
>peripheral units