3.1 General rules
719 S Titrino
61
RS receive errors:
E36 Parity
Exit:
both devices are the same.
E37 Stop
Bit
Exit:
both devices are the same.
E38
Overrun error. At least 1 character could not be read.
Exit:
E39
The internal working-off buffer of the Titrino is full (>82
characters).
Exit:
RS send errors:
E40
DSR=OFF No proper handshake for more than 1 s.
Exit:
E41
DCD=ON No proper handshake for more than 1 s.
Exit:
E42
CTS=OFF No proper handshake for more than 1 s.
Exit:
E43
The transmission of the Titrino has been interrupted with XOFF for
at least 6 s.
Exit: Send XON or
E44
The RS interface parameters are no longer the same for both
devices. Reset.
E45
The receive buffer of the Titrino contains an incomplete command
(L
F
missing). Sending from the Titrino is therefore blocked.
Exit: Send L
F
or