4 error messages – Metrohm 719 S Titrino User Manual
Page 64
3.1. General rules
719 S Titrino
60
3.1.4 Error messages
Error messages are added to the status messages and separated from them
by the sign ";".
E20
Check exchange unit.
Exit: Mount Exchange Unit (properly) or &m $S.
E21
Check electrode, short circuit.
Exit: Rectify fault or &m $S.
E22
Check electrode, break.
Exit: Rectify fault or &m $S.
E23
Division by zero.
Exit: The error message disappears on next startup or on recalculation.
E26
Manual stop.
Exit: The error message disappears on next startup.
E27
Stop V reached in SET.
Exit: The error message disappears on next startup.
E28
Wrong object call up
Exit: Send correct path for object. Start path at root.
E29
Wrong value or no value allowed.
Exit: Send correct value or call up new object.
E30
Wrong trigger, this trigger is not allowed or carrying-out of action not
possible.
Exit: Send correct trigger (exception: $D) or call up new object.
E31
Command is not possible in active status. Repeat command in inactive
status.
Exit: Send new command.
E32
Command is not possible during titration. Repeat command during the
conditioning phase or in inactive status.
Exit: Send new command.
E33
Value has been corrected automatically.
Exit: Send new command.
E34
Instrument at the end of the titration and sample data is edited; the instrument
at rest or editing during filling.
Exit: &m $S.