Metrohm 719 S Titrino User Manual
Page 86
3.2. Remote control commands
719 S Titrino
82
Parameters for CAL: Calibration at Titrino.
With "ON", there are no hold points in the calibration sequence for entries, the
first buffer is measured directly.
3.2.2.36. Mode.Parameter.Calibration.ActPulse first,
all, OFF
Parameters for CAL: Output of a pulse on the I/O line "Activate", see page
131.
3.2.2.37. UserMeth.FreeMem read
only
Memory space, available for user methods or silo lines. $Q sends the number
of free bytes, e.g. "4928".
3.2.2.38. UserMeth.Recall $G
UserMeth.Recall.Name
up to 8 ASCII characters
UserMeth.Store $G
UserMeth.Store.Name
up to 8 ASCII characters
UserMeth.Delete $G
UserMeth.Delete.Name
up to 8 ASCII characters
UserMeth.DelAll $G
Management of the internal method memory: Load, store and delete methods.
An action is performed if "$G" is sent to the corresponding node just after en-
tering the name.
Do not use blank characters before and after method name!
.DelAll: Deletes all methods in the user memory.
3.2.2.39. UserMeth.List.1.Name read
only
UserMeth.List.1.Mode read
only
UserMeth.List.1.Quantity read
only
UserMeth.List.1.Bytes read
only
UserMeth.List.1.Checksum read
only
for each method
List of the methods in the user method memory with the following
characteristics:
.Name:
Name of the method
.Mode: Mode
.Quantity: Measured
quantity
.Bytes:
Number of bytes of the user memory used by the method
.Checksum: Checksum of the method, see 3.2.2.59.
3.2.2.40. Config.PeriphUnit.CharSet
Epson, Seiko, Citizen, HP, IBM
Selection of the character set and the graphics control characters of the Ti-
trino.
IBM means the IBM character set following character set table 437 and IBM
graphics control characters. Select 'IBM' for work with the computer.
3.2.2.41. Config.PeriphUnit.Balance Sartorius,Mettler,Mettler AT,
AND,Precisa
Config.PeriphUnit.Plot
U, dU/dt, V, dV/dt, U(rel), T