Metrohm 719 S Titrino User Manual
Page 90
3.2. Remote control commands
719 S Titrino
86
Info.CalibrationData.Inp1.Date read
only
Info.CalibrationData.Inp1.ElectrodeId read
only
identical for .Inp2 and .Diff
pH calibration data for measuring input 1. After the calibration, the data are
entered automatically together with the date of the calibration and the elec-
trode identification, see 3.2.2.34.
Calibration data can be entered. They are accepted with &Info.CalibrationData
$G. If calibration data are entered, the calibration date is deleted.
3.2.2.60. Info.Checksums $G
Info.Checksums.ActualMethod read
only
The checksums can be used to identify the content of a file unequivocally,
e.g. files with identical content
have identical results of the checksums. An empty file has checksum "0". The
calculation of the checksums is triggered with $G.
.ActualMethod: Result of the checksum of the current method in the working
memory. Identical methods with different method names have the
same results of the checksum.
3.2.2.61. Info.DetermData $G
Info.DetermData.Write ON,
OFF
Info.DetermData.ExV read
only/read+write
Recalculation of the measuring data is triggered with $G.
.Write:
With "ON", the following nodes can be overwritten:
&Info.TitrResults.Var.C4X (X = 0...5) and &Mode.Name.
.ExV:
Volume of the exchange unit, with which the determination was
executed
3.2.2.62. Info.TitrResults.RS.1.Value read
only
etc., up to .9
Info.TitrResults.EP.1.V read
only
Info.TitrResults.EP.1.Meas read
only
etc., up to .2
Info.TitrResults.Var.C40 read
only/read+write
etc., up to .C47
.RS: Values of the calculated results.
.EP: Endpoints with SET:
Volume coordinate in mL, e.g. "1.2340"
Measured value coordinate in pH "5.12", mV (with U and Ipol) "-
241" or µA (with Upol) "43.7".
.Var: Various variables. You may overwrite the variables C40...C45, see
3.2.2.95.
C40: Initial measured value in pH "5.12", mV (with U and Ipol)
"41", µA (with Upol) "43.7" or °C (with T) "25.0". In MEAS final
measured value.
C41: End volume with SET in ml, "12.5360".
C42: Time from start of titration to end in s, "62".
C43: Volume drift on start of a SET titration from the conditioning
in ul/min, "3.5".