Metrohm 719 S Titrino User Manual
Page 69
3.2 Remote control commands
719 S Titrino
65
*Parameter
Tree part "Parameters for SET"
├
.
SET1
Control parameters for EP1
│ ├
.EP
Endpoint 1
depends on meas.quant. 3.2.2.10.
│ ├
.UnitEp
Unit of endpoint
read only
ditto
│ ├
.Dyn
Dynamics
depends on meas.quant. 3.2.2.11.
│ ├
.UnitDyn
Unit of dynamics
read only
ditto
│ ├
.MaxRate
Maximum dosing rate
0.01...10...150, max.
ditto
│ ├
.MinRate
Minimum dosing rate
0.01...25.0...9999 ditto
│ ├
.Stop
Titration stop
│ │ ├
.Type
Type of stop criterion
drift, time
3.2.2.12.
│ │ ├
.Drift
Stop drift
1...20...999 ditto
│ │ ├
.Time
Switch-off delay time
0...10...999, inf
ditto
│ │ ├
.StopT
Stop time
0...999 999, OFF ditto
├
.
SET2
Control parameters for EP2, as for EP1
├
.
TitrPara
Titration parameters
│ ├
.Direction
Titration direction
+, -, auto 3.2.2.13.
│ ├
.XPause
Pause 1
0…999 999
3.2.2.14.
│ ├
.Start V
Start volume
│ │ ├
.Type
Type of start volume
abs., rel., OFF 3.2.2.15.
│ │ ├
.V
Volume for absolute start volume
0...999.99 ditto
│ │ ├
.Factor
Factor for relative start volume
0...±999 999
ditto
│ │ ├
.Rate
Dispensing rate for start volume
0.01...150.0, max. ditto
│ ├
.Pause Pause
2
0...999 999
3.2.2.16.
│ ├
.ExtrT
Extraction time
0...999 999
3.2.2.17.
│ ├
.MeasInput Measuring
Input
1, 2, diff.
3.2.2.18.
│ ├
.Ipol
Polarization
current
0...1...±127 ditto
│ ├
.Upol
Polarization voltage
0...400...±1270 ditto
│ ├
.PolElectrTest
Test for polarized electrodes
ON, OFF ditto
│ ├
.Temp
Titration temperature
-170.0...25.0...500.0 3.2.2.19.
│
├
.
StopCond
Stop conditions
│ ├
.VStop
Stop volume
│ │ ├
.Type
Type of stop volume
abs., rel., OFF
3.2.2.20.
│ │ ├
.V
Volume for absolute stop volume
0...99.99...9999.99 ditto
│ │ ├
.Factor
Factor for relative stop volume
0...±999 999 ditto
│ ├
.FillRate Filling
rate
0.01...150.0,
max. 3.2.2.21.
│
├
.
Statistics
Statistics
│ ├
.Status
Status of statistics calculation
ON, OFF 3.2.2.22.
│ ├
.MeanN
No. of individual determinations
2...20 ditto
│ ├
.ResTab Result
table
│ │ ├
.Select
original,delete n,delete all
ditto
│ │ ├
.DelN
Deletion of individual results
1...20 ditto
│
├
.
Presel
Preselections
│ ├
.Cond
Conditioning
ON, OFF 3.2.2.23.
│ ├
.DriftDisp
Display of drift during cond.
ON, OFF
ditto
│ ├
.DCor
Drift Correction
│ │ ├
.Type
Drift type
auto, man., OFF 3.2.2.24.
│ │ ├
.Value
Value for manual drift correction
0.0…99.9 ditto
│ ├
.IReq
Request of Id's after start
id1, id1&2, all, OFF 3.2.2.25.
│ ├
.SReq
Request of smpl size after start
value, unit, all, OFF ditto
│ ├
.ActPulse
Output of a pulse
first, all, cond., OFF 3.2.2.26.