beautypg.com

Metrohm 719 S Titrino User Manual

Page 17

background image

2.4

Tutorial

719 S Titrino

13

, or type

a text as unit, see page 8.

The Titrino prompts for the calculation of the next result.

Quit the formula entry by pressing twice.

In order to be able to calculate the result, enter the calculation constants used in the for-

mula.

Entry of the calculation constants

0.1

36.47

• Press .

The constants which have been used in the formula

are requested: input with digit keys, confirm with

.

C01: Concentration of your titrant= 0.1 mol/L.

Enter 0.1.

C02: Molar mass of HCl = 36.47 g/mol

Enter 36.47.

Now only the constant C00 is missing to perform the calculation. This constant is re-

served for the sample size. It can be entered directly.

RS1=EP1*C01*C02/C00

>formula

RS1 text

RS1

>formula

RS1 decimal places

2

>formula

RS1 unit

%

RS?

C-fmla

>C01 0.0

C-fmla

>C02 0.0